목적 : To evaluate the glaucoma-diagnostic ability of the ganglion cell-inner plexiform layer (GCIPL) thickness difference across the temporal raphe in highly myopic eyes.
방법 : Using a Matlab-based customized program (the GCIPL Hemifield Test), a positive test result was automatically declared if the following two conditions were met: (1) the horizontal line is detected for longer than one-half of the distance from the temporal inner elliptical annulus to the outer elliptical annulus, and (2) the average GCIPL thickness difference within 10 pixels of the reference line, both above and below, is 5 µm or more. The glaucoma-diagnostic ability was computed using the area under the receiver operating characteristic curve (AUC).
결과 : We consecutively enrolled a total of 195 highly myopic eyes (axial length [AL] >26.5mm) of 195 subjects: 93 glaucoma patients along with and 102 non-glaucomatous subjects. Among the glaucomatous eyes, GCIPL Hemifield Test positivity was shown in 92.5% (86 of 93), significantly higher than that for the non-glaucomatous eyes (4.90%, 5 of 102; P <0.001). The AUC value for the GCIPL Hemifield Test was excellent (0.938; sensitivity 92.50%, specificity 95.10%) and was the best compared with those for any of OCT parameters.
결론 : In highly myopic eyes, determination of the presence or absence of GCIPL thickness difference across the temporal raphe via OCT macula scan can be a useful means of distinguishing the glaucomatous damage.
|